Photometry subsystem add-ons
The IM-10 is a single channel photometry add-on to your imaging system. It is ideal to measure very fast or ultra low light levels. Photometry subsystem includes photometer with high sensitivity analog or photon counting detection; electronics and FeliX32™ software package and BryteBox interface. May require IM-9 dual C port adaptor, in order to accommodate both photometer and camera.
Back to top.
Single Channel Photometer
High sensitivity PMT detector:
- Photon counting mode: for low signal, as in most ratio fluorescence measurement.
- Analog modes: for high signal
- Detector wavelengths: UV- NIR with appropriate PMT
Precise field-of-view control
- Precise field-of-view control
- A viewing eyepiece for Parfocal viewing.
An adjustable aperture for selecting the area of interest.
- Easy to switch from View to Measure via a flipping mirror.
Connect to the microscope on any of the available ports via a C-mount adaptor.
Back to top.
Dual Channel Photometer
Has all the properties of a single channel photometer, plus: Allows for simultaneous detection of two emission wavelengths by means of two independent PMT detectors. Has provision for beam splitter and dichroic cube for emission wavelength selection. Dichroic assembly is placed within the photometer for selection of the two emission wavelength ranges. Provides high-speed (millisecond) detection for emission-shifted probes. Data acquisition rates of up to 1000 ratios per second are possible. Back to top.
Dual C-Port Adaptor with Flipping Mirror
The IM-9 is a Dual C port adaptor for microscope with flipping mirror allowing user to select between two different emission detection devices, for example camera and photometer. Primarily used to mount two optical detectors one output for photometer, another output for camera. Both ports are identical. The flipping mirror allows easy and rapid switch between the two detectors without a hardware change. Good for sequential, not simultaneous, measurement. Very useful to retrofit old fluorescence scopes. Attaches directly to the C-mount on a microscope. Provides one input and two standard C-Port outputs. Back to top.
Fluorescence Lifetime Upgrades
Pulsed laser and flash lamp excitation sources and detectors can be added to your microscope to measure Fluorescence and phosphorescence lifetimes. FeliX32™ advanced analysis software is used for curve fitting the decay data. Dual View Image splitter module The IM-5B adds popular dual Image splitter modules to your EasyRatioPro system. This allows for emission ratio and FRET Imaging applications like Indo-1, JC-1 membrane potential or CFP/YFP FRET.
Back to top.
Dual View Image Splitter Module
The IM-5B adds popular dual Image splitter modules to your EasyRatioPro system. This allows for emission ratio and FRET Imaging applications like Indo-1, JC-1 membrane potential or CFP/YFP FRET. Back to top.
|