System Enhancements


Phosphorescence and Steady State (T1a)

PTI

Adds phosphorescence lifetime and steady state capability to systems, includes pulsed xenon source with QuadraScopic, computer controlled, monochromator for excitation, coupled to the second entrance port of sample compartment; added to the sample compartment second excitation channel optics; for detection a gated detector with 1527 PMT is coupled to the second exit port of the emission monochromator, one can select detectors by means of a flipping mirror.

The T-1A requires an open sample compartment port since it is directly coupled to the sample compartment and an open port on the current system's emission monochromator.

T1a Monchromator Specifications

Focal length

200 mm

Aperture ratio

f/4 (calculated using grating width)

Wavelength range

180 nm to 24 microns

Bandpass

continuously adjustable from 0 to 20 nm

Reciprocal linear dispersion

4 nm/mm

Resolution

0.5 nm

Throughput

60% at 300 nm

Scattered light

0.02% two bandwidths from 365 nm Hg line

Accuracy

+/- 1 nm (using manual wavelength control)

Reproducibility

+/- 1 nm

Optical path height

76 mm

Grating size

50 x 50 mm

 

T1a Detector Specifications

Detector

Coming soon

 

T1a Lamp Specifications

Lamp type

Pulsed xenon source

Wavelength range

From 200 to 2,000 nanometers

Pulse repetition rate

From 1 to 300 pulses per second

Pulse width

Approximately 2 microseconds

Back to top.

 

 

Steady State (T2)

PTI

For steady state measurement capability to lifetime systems, includes high intensity continuous xenon source with QuadraScopic, computer controlled, monochromator for excitation, coupled to the second entrance port of sample compartment; added to the sample compartment second excitation channel optics; for detection there is a high sensitivity photon counting/analog detector with 1527 PMT added to second exit port of emission monochromator, one can then choose between detectors by means of a flipping mirror.

The T-2 requires an open sample compartment port since it is directly coupled to the sample compartment and a open port on the current system's emission monochromator.

T2 Detector Specifications

Input

+/- 15 VDC, 250 mA

High voltage

-200 to -1,100 VDC manually adjustable (LCD display indicates actual cathode voltage)

External high voltage adjust

0 to +5 VDC (0 = -200 V, 5 = -1,100 V), continuously adjustable

Input regulation

+/- 0.05% max. (for 15 V +/- 1 V input)

Load regulation +/- 0.05% max.
Ripple 100 mV p-p max.
Temperature coefficient +/- 0.0% max. (+5 to +40 deg. C)
Drift +/- 0.03%/hr max. (after 15 minute warm-up)

 

T2 Monochromator Specifications

Focal length

200 mm

Aperture ratio

f/4 (calculated using grating width)

Wavelength range

180 nm to 24 microns

Bandpass

continuously adjustable from 0 to 20 nm

Reciprocal linear dispersion

4 nm/mm

Resolution

0.5 nm

Throughput

60% at 300 nm

Scattered light

0.02% two bandwidths from 365 nm Hg line

Accuracy

+/- 1 nm (using manual wavelength control)

Reproducibility

+/- 1 nm

Optical path height

76 mm

Grating size

50 x 50 mm

 

T2 PMT Specifications

Spectral response

185 to 680 nm

Cathode sensitivity
      Luminous
     Radiant

60 µA/lm
400 nm–60 mA/W

Anode sensitivity (at 1000 V)      Luminous
     Radiant

400 A/lm
400 nm–4.0 x 105 A/W

Low dark current

0.1 nA

Low dark counts

(R1527P) - 10 cps

 

T2 Lamp Specifications

Lamp power capacity

75 to 200 watts

Height

100 mm (3.9 inches)

Width

100 mm (3.9 inches)

Length

210 mm (8.3 inches)

Weight

1.9 kg (4.2 pounds)

Window diameter (D)

65 mm

Back to top.

 

 

Extended Phosphorescence Detection (T3)

PTI

Extended phosphorescence detection option for fluorescence lifetime systems (for lifetimes beyond 500 nanoseconds), adds gated detector with 1527 PMT to second exit channel of emission monochromator, detector selection by means of flipping mirror.

T3 Detector Specifications

Detector

Coming soon

Back to top.

 

 

Dual Emission (T5)

PTI

Converts TM-3 lifetime system to L-format LaserStrobe™ system. Adds second emission monochromator, electrometer and 1527 PMT, sample compartment optics, filter holder and mechanical shutter.

T5 Emission Monchromator Specifications

Wavelength range

180 nm to 24 microns, continuously tunable (useful detection wavelength range dependent on grating and photomultiplier tube)

Bandwidth

0 to 25 nm, continuously tunable

Resolution

0.5 nm

Accuracy

+/- 0.5 nm

Reciprocal linear dispersion

4 nm/mm

Throughput

60% at 400 nm

Stray Light < 0.02% two bandwidths from 365 nm Hg line

 

T5 Stroboscopic Detector Specifications

Wavelength range

185 to 680 nm with standard R1527 PMT