System Specifications
 |
Capable of measuring lifetimes of 7pM fluorescein
Range of Measurable Lifetimes: 100 picoseconds to ~20 microseconds (sample intensity dependent) |
| Detection Limit |
Sensitivity allows for the measurement of lifetimes of 7 pM fluorescein |
|
| Minimum Measurable Lifetime |
< 100 ps |
|
| Data Acquisition Rate |
< 30 seconds typ. |
|
| Excitation Range |
337 to 990 nm (depending on dye used)
235 to 345 nm (with optional frequency doubler) |
|
| Emission Range |
185 to 680 nm (optional to 900 nm) |
|
| Pulse Width |
800 picoseconds |
|
| Peak Power |
275 kW at 500 nm, 5 Hz |
|
| Peak Energy |
220 microjoules per pulse at 500 nm, 5 Hz 275 kW at 500 nm |
|
| Repetition Rate |
2 to 20 Hz |
|
| Detection |
Time domain, patented stroboscopic technique |
|
| Bandpass |
0.04 nm |
|
| System Control |
Computer interface with spectroscopy software |
|