FluoDia T70 Hardware Specifications
Light source |
Quartz Halogen lamp |
|
Light guide |
Quartz fiber + condenser lens |
|
Optics |
Top-to-top |
|
Interference filter |
Filter ports: 4
Automatic switching |
|
Detector |
PMT |
|
Measurement method |
Photon counting |
|
Wavelength range |
320–830 nm |
|
Integration time |
10 msec–10sec |
|
Measurement time |
~ 40 sec (96 wells) |
|
Number of averages |
1–100 times/well |
|
Sample mixing |
Linear and butterfly shaking |
|
Temperature control |
Exclusive heating stage |
|
Temperature range |
RT +5°C–75°C |
|
Plate type |
6-384 well plate, 1536 wells (Option),
Transparent, black and white |
|
| PC laptop (option) |
|
FP
The FP is a highly sensitive fluorescence plate reader. It can accommodate up to four sets of filters for excitation and emission wavelength selection. Included are one set of filters (Exc. 486 ± 10 nm , Emis. 530 ± 10 nm for fluorescein) and a user friendly software. FP can accommodate 6, 12, 24, 48, 96, 384 well (1536 well optional) microplates of any color, shape, made of any material. It is the number one choice for customers who need a highly sensitive platereader without temperature control. Computer is required but not included.
Back to top.
FP-1
FP-1 is a highly sensitive high temperature fluorescence plate reader. It can accommodate up to four sets of filters for excitation and emission wavelength selection. Included are one set of filters (Exc. 486 ± 10 nm , Emis. 530 ± 10 nm for fluorescein), a flat heater plate with a temperature limit of 75°C and a user friendly software. FP-1 can accommodate 6, 12, 24, 48, 96, 384 well (1536 well optional) microplates of any color, shape, made of any material. It is the number one choice for customers who need a platereader that can accommodate microplates of different types. Computer is required but not included.
Back to top.
FP-2
FP-2 is a high sensitivity high temperature fluorescence plate reader specially designed for exceptional accurate temperature control of the samples in a 96 well microplate. It can accommodate up to four sets of filters for excitation and emission wavelength selection. Included are one set of filters (Exc. 486 ± 10nm , Emis. 530 ± 10 nm for fluorescein), a 96 well heater plate with a temperature limit of 75°C and a user friendly software. FP-2 is specifically designed for U or V-shaped 96 well microplate. It can also accommodate strip tubes of 8 wells or 12 wells. With accuracy of ± 0.3°C, homogeneity of ± 0.3°C across the plate, this is a model for 96 well microplate users who wants very accurate temperature control and for strip tube users looking for a microplate reader. Computer is required but not included.
Back to top.
FP-3
FP-3 is a highly sensitivity high temperature fluorescence plate reader specially designed for exceptional accurate temperature control of the samples in a 384 well microplate. It can accommodate up to four sets of filters for excitation and emission wavelength selection. Included are one set of filters (Exc. 486 ± 10 nm , Emis. 530 ± 10 nm for fluorescein), a 384 well heater plate with a temperature limit of 75°C and a user friendly software. FP-3 is specifically designed for U or V-shaped 384 well microplate. With accuracy of ± 0.3°C and homogeneity of ± 0.3°C across the plate, this is a model for 384 well microplate users who requires very accurate temperature control. Computer is required but not included.
Back to top.
FP-Z
FP-Z has all the features of FP. In addition, it allows the user to manually set the focal point for measurement. This is extremely useful for customers who want their measurement to be "focused", for example, on the cells that are attached on the bottom of a plate, which used to require a bottom reading platereader. FP-Z can perform this type of application with sensitivity that is thousands times greater than a bottom reading platereader.
Back to top.
FP-1Z
FP-1Z has all the features of FP-1. In addition, it allows the user to manually set the focal point for measurement. This is extremely useful for customers who want their measurement to be "focused", for example, on the cells that are attached on the bottom of a plate, which used to require a bottom reading platereader. FP-1Z can perform this type of application with sensitivity that is thousands times greater than a bottom reading platereader.
Back to top.
FP-2Z
FP-2Z has all the features of FP-2. In addition, it allows the user to manually set the focal point for measurement. This is extremely useful for customers who want their measurement to be "focused", for example, on the cells that are attached on the bottom of a 96 well plate, which used to require a bottom reading platereader. FP-2Z can perform this type of application with sensitivity that is thousands times greater than a bottom reading platereader.
Back to top.
FP-3Z
FP-3Z has all the features of FP-3. In addition, it allows the user to manually set the focal point for measurement. This is extremely useful for customers who want their measurement to be "focused", for example, on the cells that are attached on the bottom of a 384 well plate, which used to require a bottom reading platereader. FP-3Z can perform this type of application with sensitivity that is thousands times greater than a bottom reading platereader.
Back to top.
|