PTI, Photon Technology International, The Fluorescence Solutions Company

 

or

 
   

The Fluorescence Solutions Company

 
 

RatioMaster™
Fluorescence Microscopy with Photo Uncaging

 
  Fixed Spot Illumination
 
 
Fluorescence Microscopy quote
 
 
 

Fixed Spot Illumination System SI-1

The standard Fixed Spot Illumination System includes a Laser light source, and a Spot Illumination module comprised of optics, dichroic mirror and a light guide. Light Source Options are available.

 

Laser

The Laser is a 355 nm laser with 30 mW average power. .

Flash Lamp for Photo Uncaging with the Easy Ratio Pro

Specifications

Wavelength 355 nm
Beam Diameter < 400 µm
Average Power 30 mW
Energy Per Pulse
3 µJ
Rep Rate Up to 100 kHz
Coupling Fiber Optically

 

Spot Illumination Module

Spot illumination module consists of optics, dichroic mirror and a light guide.

The spot illumination module allows spot illumination of a spot in the center of the field of view (and simultaneous full field illumination, except when ordering it with a 100% mirror). Spot size depends on diameter of light guide, magnification of objective and microscope. When ordering please specify approximate spot size requested and magnification of objective. Please make sure that objectives, optics and filters of the microscope transmit/reflect the desired wavelengths. Most long distance IR objectives cannot be used for any spot illumination.

We strongly recommend not combining a flash lamp system with a
spot illumination module. For applications requiring a rather small spot a light guide with a small diameter is needed and only a small amount of light can be coupled into a light guide with a small diameter. The flash lamp is not a point source and emits light in almost 360° resulting in a light output that is not sufficient for the majority of experiments.

Photo Uncaging Flash Lamp System JML-C@

 

 