Model A-500 Lamp Housing

Also applicable to the MilliArc

CONTENTS

1

DESCRIPTION

Introduction

Power Supply

Lamps

2

INSTALLATION

Installing the Lamp

Connections

Focusing

3

OPERATION

Arc Lamps

Tungsten-Halogen Lamps

 

 

 

DESCRIPTION

Introduction

The PTI Model A-500 Arc Lamp Housing is designed to accommodate a variety of Xenon, Mercury and Mercury-Xenon high pressure short arc lamps with power ratings of 75 to 200 watts, as well as vertical-filament technical Tungsten-Halogen lamps. The housing features an adjustable spherical reflector for increased light out- put and a lens holder that accommodates a variety of focus and collimation lenses. The unit is designed for efficient ozone control.

WARNING: Never look directly at an operating arc lamp; severe eye injury will result. Wear U.V. protective lenses, such as a welder's helmet, when working around operating arc lamps.

 

Return to contents.

Power Supply

The A-500 is compatible with PTI Models LPS-220 and LPS-250 power supplies, but may be used with power supplies from other manufacturers. In any case, please refer to the power supply manual for electrical connections and operating instructions.

WARNING: Use caution when disconnecting the power supply from the lamp housing. High voltage arcing can occur between the power lead connectors and any nearby metallic objects, and can cause personal injury.

 

Return to contents.

Lamps

Compact Arc Lamps

High pressure gas discharge lamps having an arc length which is small compared with the size of the electrodes are called short arc or compact arc lamps. Depending on rated wattage and intended application, the arc length of these lamps may vary from about a third of a millimeter to about a centimeter. These lamps have the highest luminance and radiance of any continuously operating light source and are the closest approach to a true "point" source.

The envelope is made from optically clear quartz material of various grades and has a spherical or ellipsoidal shape. The grade of the quartz will determine the amount of ozone generated. The most widely used material for the electrodes is tungsten.

Most compact arc lamps are designed for DC operation. This results in better arc stability and substantially longer life. DC systems consist of an igniter and a regulated power supply. High voltage pulses (up to 50,000 volts) break down the gap between the electrodes, ionize the gas and heat the cathode tip to thermionic emitting temperatures.

Xenon Lamps

Xenon compact arc lamps are filled with several atmospheres of xenon gas. They reach 80% of final output within 10 minutes or less of starting. The arc color is very close to daylight (6000 deg. K). The spectrum is continuous in the visible range and extends far into the ultraviolet. A Xenon lamp exhibits strong lines in the near infrared between 800 and 1000 nm and some weak lines in the blue portion of the spectrum.

Xenon compact arc lamps are made with rated wattages from 75 to 30,000 watts and are available for operation in either a vertical or horizontal position. The breakdown voltage between the electrodes will run from 10kv for a small lamp up to 60kv or more for lamps rated 30kw.

The luminous efficacy of Xenon compact arc lamps is approx. 30 lumens per watt at 1000 watts, 45 lumens per watt at 5000 watts, and over 150 lumens per watt at 20kw.

Mercury-Xenon Lamps

A Mercury-Xenon lamp contains a specific amount of mercury and a small amount of xenon added at a pressure exceeding one atmosphere. The xenon is necessary to facilitate starting and to sustain the arc until the mercury is fully vaporized; it also reduces the warm-up period. Normal warm-up time is 10-15 minutes.

Mercury lamps are sensitive to cooling because the bulb temperature determines the vapor pressure. The lamp can be over-cooled to the point that full output in the mercury spectrum is never achieved. The cooling water should be ordinary tap water. Chilled water may decrease the operating voltage and interfere with the proper evaporation of mercury. In some cases, the mercury may not evaporate at all, causing unsuitable performance and shortened lamp life.

Typical steady state voltage of a Mercury-Xenon lamp is higher than that of a xenon lamp. The output in the visible range consists mainly of four mercury lines and some continuum, due to the high operating pressure. A properly warmed lamp will show no significant trace of the xenon gas spectrum.

Mercury-Xenon lamps are available in wattages from 200 to 7000 watts. The luminous efficacy is approximately 50 lumens per watt at 1000 watts and about 55 lumens per watt at 5000 watts.

Tungsten Lamps

Technical lamps consist of a coiled tungsten filament mounted in a precision glass envelope. The envelope may have a vacuum or, more commonly, be filled with an inert gas such as argon or krypton. Typical technical lamp operating parameters are 2.5 to 12 volts and .02 to 1 amp. Color temperature ranges from 2,200 to 3,000 degrees Kelvin; lamp life may be as high as 30,000 hours.

Tungsten-Halogen lamps feature a tungsten coil filament mounted in a quartz glass envelope that has been filled with an inert gas plus a trace of halogen (normally bromine). This gas creates the “halogen cycle”: tungsten that has evaporated from the filament combines with the halogen gas. Convection currents within the bulb carry this gas to the quartz wall where it is cooled and then returned to the proximity of the filament. The heat of the filament causes the tungsten and bromine to separate, and the tungsten is then deposited on the cold portion of the filament.

This regenerative process prolongs the life of the filament considerably, and also eliminates blackening of the bulb by preventing the evaporated tungsten from condensing on the envelope. The Halogen lamp color temperature runs from 2900 to 3400 deg. Kelvin and are available in wattages from 10 to 250 at operating voltages from 6 to 24; lamp life ranges from 10 to 2500 hours. Luminous efficiency is approximately 22 lumens per watt.

Tungsten-Halogen lamps must be operated at voltages thatmaintain an envelope temperature between 250 and 350 deg. C. Cooler temperatures will not allow the halogen cycle to take place, thus causing bulb blackening and shorter life; higher temperatures will cause oxidation of the conductors and lead to premature lamp failure.

Recommended Lamps

The following lamps are recommended for use in the A-500, most of which are available from PTI.

Osram XBO 75W/Z Xe
Osram XBO 150W W/S Xe
Ushio UXL 75W Xe
Ushio UXL 75W CLS Xe
Ushio UXL 100D Xe Hg
Ushio UXL 200H Hg Xe
Ushio UXL 151H Xe
Gilway L7412 75W T/H

NOTE: Lamps purchased directly from other manufacturers may not meet our specifications for operation in the A-500 and may cause permanent damage to the housing. Use of such lamps will void the Warranty.

Note that higher wattage lamps do not necessarily yield more light intensity. When higher illumination intensity is needed, lamps must be selected with greater brightness, and this does not always increase with lamp wattage.

Lamp Handling

WARNING: Compact arc lamps contain highly pressurized gas, and present an explosion hazard even when cold. Wear face protection, such as a welder's helmet, whenever handling lamps.

 

Special storage cases are provided to eliminate possible hazards during shipping and handling. Safety goggles and soft cotton gloves should be worn when removing and installing lamps. Never touch the quartz envelope with bare hands; such handling may lead to deterioration and premature failure. If accidentally handled, clean the lamp surface with an alcohol swab to remove any residue.

WARNING: Never look directly at an operating arc lamp; severe eye injury will result. Wear U.V. protective lenses, such as a welder's helmet, when working around operating arc lamps.

 

Polarization

Some lamps can only be mounted one way in the A-500 housing since the anode (+) and cathode (-) have different diameters, thus making accidental polarization reversal nearly impossible. However, some lamps have the same diameter anode and cathode, allowing room for error. Refer to the lamp manufacturer's data sheet for proper identification of the anode and cathode.

IMPORTANT: Reversed polarization will result in immediate and permanent damage to the lamp electrodes. PTI will not warrant or replace a lamp which has been improperly mounted.

A lamp that has been fired with reversed polarization will have obvious physical damage to the electrodes. A damaged lamp will fire, but it will exhibit unstable performance and a severely shortened operating life.

Lamp Stability

Short term stability is measured over seconds, while long term stability is measured over minutes, hours, or even days.

Short term stability is affected by arc "wander," "flare" and "flutter." Arc wander is the movement of the attachment point of the arc on the cathode surface. Typically the arc moves around the conical cathode tip in a circular fashion, taking several seconds to move a full circle. Arc flare refers to the momentary change in brightness as the arc moves to an area on the cathode having a preferential emissive quality over the previous attachment point. Arc flutter is the rapid side-to-side displacement of the arc column as it is buffetted by convection currents in the xenon gas which are caused as the gas is heated by the arc and cooled by the envelope walls.

Arc wander and flare can be reduced by a slight decrease in the operating current. For example, a 75 watt xenon lamp rated at 5.4 amps may be operated at 4.5 amps for the first one or two minutes of operation, after which the current should be brought up to the specified normal operating level.

Lamp Life

The useful life of compact arc lamps is determined primarily by the decrease of luminous flux caused by the deposit of evaporated electrode material on the inner wall of the envelope. Frequent ignition accelerates electrode wear and hastens the blackening of the envelope. Average lamp life is based on approximately 20 minutes of operation for each ignition. The end of the lamp life is the point at which the UV output has decreased by approximately 25%, the arc instability has increased beyond 10%, or the lamp has ceased to operate under specified conditions. Lamps should be replaced when the average lamp life has been exceeded by 25%.

As the lamp ages, the operating voltage will increase. Lamp current should be decreased to maintain output until the minimum operating current is reached. At this time the lamp should be replaced.

Lamp life varies with different types. Check the manufacturer's specifications for the rated lamp life.

Return to Contents

 

 

 

INSTALLATION

Lamp Installation

The arc lamp, if ordered with the lamp housing, is shipped separately in its original packaging and must be installed before using the A-500.

WARNING: Compact arc lamps contain highly pressurized gas, and present an explosion hazard even when cold. Wear face protection, such as a welder's helmet, whenever handling lamps. When installing or replacing the lamp, always use the protective sleeve, if provided, or wear soft cotton gloves to avoid touching the envelope. Whenever servicing the lamp, always disconnect all electrical connections to the lamp housing before opening it.

 

  1. Remove the six phillips-head screws on the lamp housing cover, and slide the lid straight up.

  2. If it has not already been installed, insert the lamp cathode adapter (if required) into the lamp mounting base and tighten the set screw.

  3. Insert the lamp, cathode (-) down, into the cathode adapter or the lamp mounting base and tighten the set screw.

  4. Slip the anode connector over the lamp anode (+) and tighten the set screw. Ensure all electrical connections are secure. Refer to the figures below for the 75 watt and 100 watt arc lamps.

  5. Replace the lid and six lid screws.

For vertical-filament Tungsten-Halogen lamps, a special lamp socket is installed in the lamp mounting base. If the socket is not already installed, insert the socket and align the lamp lead holes such that they are perpendicular to the beam path of the lamp housing. Tighten the set screw. Insert the lamp into the socket and replace the lid. Refer to the figure below.

Return to Contents

Connections

WARNING: Be sure the power supply is turned off before making electrical connections.

 

For arc lamps, connect the igniter to the power supply. Connect the white lead from the igniter to the A-500 lamp housing red (+) connector. Connect the black lead from the igniter to the black (-) connector.

For Tungsten-Halogen lamps, do not use the igniter. Connect the red lead from the power supply to the A-500 lamp housing red (+) connector. Connect the black lead from the power supply to the black (-) connector.

Return to Contents

Focusing

Since no two arc lamps are identical and since it is impossible at best to perform a lamp replacement without altering any adjustments, the lamp must be refocused after replacement.

WARNING: Wear U.V. protective lenses, such as a welder's helmet, when focusing the lamp.

 

  1. Set all lamp focus adjustments and lens(es) to their midpoints.

  2. Using the lamp adjusting screws, align the lamp to achieve best centering, then best focus. Refer to the figure below.

  3. Adjust the lens to focus the primary image.

  4. Using the reflector adjusting screws, adjust the reflector to focus the secondary image onto the primary image.

Return to Contents

 

 

 

OPERATION

Arc Lamps

Compact arc lamps should not be operated below 85% or in excess of 110% of rated current. Operation below rated current results in arc instability and reduced lamp life. Always allow 10-15 minutes for proper warm-up of Xenon and Mercury-Xenon lamps.

  1. Set the DISPLAY selector on the power supply to AMPS.

  2. Turn the power supply on.

  3. Set the CURRENT to approximately 5 to 5.4 amps for 75 Watt Xenon lamps, 7.5 amps for 150 Watt Xenon lamps.

  4. Press and hold the IGNITE button until the lamp ignites.
  5. After the lamp is ignited, set the DISPLAY selector to WATTS and adjust the CURRENT control to provide the lamp with its rated power.

  6. The lamp is extinguished by turning the power supply off.

Compact arc lamps should not be operated below 85% or in excess of 110% of rated current. Operation below rated current results in arc instability and reduced lamp life. Always allow 10-15 minutes for proper warm-up of Xenon and Mercury-Xenon lamps.

Return to Contents

Tungsten-Halogen Lamps

  1. Set the DISPLAY selector on the power supply to WATTS.

  2. Set the CURRENT to 0 (MIN).

  3. Turn the power supply on.

  4. Slowly turn the current up, making sure not to go over the rated wattage of the lamp.

  5. The lamp is extinguished by turning the power supply off.

Return to Contents

 

 