PTI, Photon Technology International, The Fluorescence Solutions Company

 

or

 
   

The Fluorescence Solutions Company

 
 

ImHD
High Definition Fluorescence Imaging

 
  System Specifications
 
Fluorescence Imaging
 
Imaging system quote
 
ImHD Flyer
Fluorescence Imaging
 

System Specifications

Excitation Wavelength Range 290–650 nm
Wavelength Selection Speed < 2 milliseconds point-to-point
Data Acquisition Rate 100 fps at 2560 x 2160 to 900 fps at 320 x 240
Light Source High power continuous Xenon arc lamp
Repetition Rate 1 to 200 Hz ( software controllable)
Focal Length 150 nm
Bandpass 0–25 nm, continuously adjustable (computer control available)
Wavelength Accuracy +/- 1 nm
Resolution 0.5 nm
Light Delivery 2 m Liquid Light Guide
System Control Warp Drive Imaging™ interface

HDI Camera Specifications

Resolution of 2560 X 2160 pixels 2560 x 2160
Pixel Size um2 6.5 X 6.5
Quantum Efficiency % @ 590 nm Peak 57 %
Dynamic Range A/D 16 bits
Readout Noise @ 30 fps / 100 fps e-RMS <1.4 / < 2
Imaging Frequency, frame rate 30
Exposure Time s. Rolling Shutter/ Global Shutter 500 µs to 2 Sec
Region of Interest User
Optical Input F-Mount(std), C-Mount (opt)
Data Interface Dual Channel Full Camera Link

DeltaRAM X™ Specifications

Excitation wavelength range 330–650 nm
Optimum scanning range 290–610 nm
Wavelength selection speed < 2 milliseconds point-to-point
Resolution < 1 nm
Accuracy + 3/-1 nm
Wavelength bandwidth Adjustable from 0–24 nm
Stray light rejection > 104
Light delivery 0.5 meter quartz fiber optic bundle
Liquid light guide 2 meter length 2 mm core diameter
Microscope adapter User specified

 

 