| Detection Technique |
Single-Shot Transient Digitizer (SSTD) |
|
| Signal to Noise Ratio |
3000:1 or better |
|
| Minimum Measurable Lifetime |
400 ns |
|
| Data Acquisition Rate |
Up to 300 points/sec |
|
| Emission Range |
185 to 680 nm (optional to 900 nm) |
|
| Light Source |
High power Xenon flash lamp |
|
| Repetition Rate |
1 to 300 Hz (software controllable) |
|
| Pulse Width |
1–2 µs |
|
| Monochromators |
Czerny-Turner design |
|
| Focal Length |
200 mm |
|
| Excitation Grating |
1200 line/mm 300 nm blaze |
|
| Emission Grating |
1200 line/mm 400 nm blaze |
|
| Optional Grating |
75–2400 line/mm and holographic models available |
|
| Bandpass |
0–24 nm, continuously adjustable (computer control available) |
|
| Wavelength Accuracy |
+/- 0.5 nm |
|
| Wavelength Resolution |
0.06 nm |
|
| Temporal Resolution |
1 µs |
|
| System Control |
Computer in interface with spectroscopy software |
|
| Dimensions |
32 x 26 inches |